Journal of Intelligent Material Systems and Structures

 

Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Register here to gain access to SAGE's 500+ Journals Online

SAGETRACK

Sign In to gain access to subscriptions and/or personal tools.
This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
1045389X07081319v1
19/7/815    most recent
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Google Scholar
Right arrow Articles by Dugnani, R.
Right arrow Articles by Chang, F.K.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
This version was published on July 1, 2008
Journal of Intelligent Material Systems and Structures, Vol. 19, No. 7, 815-826 (2008)
DOI: 10.1177/1045389X07081319

In Vitro Atherosclerotic Plaque Characterization by Acoustic Impedance Monitoring, Part I: Sensor Modeling, Design, and Fabrication

R. Dugnani

Department of Aeronautics and Astronautics, Stanford University, Stanford, CA 94305, USA, rdugnani{at}stanfordalumni.org

F.K. Chang

Department of Aeronautics and Astronautics, Stanford University, Stanford, CA 94305, USA

This investigation aims at developing a technique to diagnose the presence of plaque in arteries by measuring the variations of the electromechanical impedance of a piezoelectric (PZT) sensor in contact with the arterial surface. The proposed technique makes use of either one or multiple PZT sensors integrated on a balloon and used prior to percutaneous transmural angioplasty interventions. The system we envision consists of two components: the active, miniaturized PZT sensors (surface-mounted on an angioplasty balloon), and the computer and software. In the proposed technique, the angioplasty balloon with the wall-mounted sensor(s) is inflated in the artery to make contact with the plaque. Initially, low pressure is used not to disrupt the plaque. The diagnostic test feeds an electrical signal that excites the sensors; by analyzing the sensors' response, information on features and properties of the plaque is obtained. Doctors will use this information to recognize unstable plaque prior to full pressurization of the angioplasty balloon. The first part of the article describes the procedure used in designing a sensor for the detection of unstable plaque by the impedance method. Both finite element methods and an analytical model — a modified version of the Krimholtz—Leedom—Matthaei model (Krimholtz et al., 1970) — are utilized in the design of the sensor. An appropriate fabrication technique for manufacturing small-scale PZT sensors is also proposed for use in smaller arteries such as the coronary arteries. Experimentation and verification of the proposed technique is presented in the second paper stemmed from this investigation.

Key Words: angioplasty • plaque detection • unstable plaque • percutaneous transmural angioplasty (PTA) • acoustic impedance • atherosclerosis detection • diagnostics technique • piezoelectric sensors • PZT.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?